117
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. Application: Testing of residual corrosion inhibitors (Oxygen scavengers) in boiler feed
water or condensate.
2. Before using clean the vials with Hydrochloric acid (approx. 20%). Rinse thoroughly with
deionized water.
3. Ideally temperature for full colour development is 25°C ± 3 °C.
4. Volume should always be metered by using suitable pipette (class A).
5. Keep blank and sample dark during colour development time. UV-light (sunlight) causes
too high measurement results.
6. Interferences:
• Iron (II) interferes at all concentrations:
Repeat the test procedure but without adding the VARIO DEHA Rgt 2 solution. If the
displayed result is above 20 μg/l subtract this value from the DEHA test result.
• Substances which reduce Iron (III) interfere. Substances which complex iron strongly
may interfere also.
• Substances who may interfere when present in concentrations at:
Borate (as Na
2
B
4
O
7
)
500 mg/l
Cobalt
0.025 mg/l
Copper
8.0 mg/l
Hardness (as CaCO
3
)
1000 mg/l
Lignosulfonates
0.05 mg/l
Manganese
0.8 mg/l
Molybdenum
80 mg/l
Nickel
0.8 mg/l
Phosphate
10 mg/l
Phosphonates
10 mg/l
Sulfate
1000 mg/l
Zinc
50 mg/l
7. There is an option to change the unit from mg/l to μg/l.
The unit mg/l is rounded, e.g.: 25 μg/l = 0.025 mg/l display 0.03 mg/l.
mg/l
μg/l
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...