33
SP600 Spectrophotometer 01/2008
1.1 Methods
Sample preparation: Reaction times must
be exactly kept !
1. Prepare the
dry
reaction apparatus (note 4) and place
it under a fume hood (toxic fumes!).
2. Pipette
50 ml of water sample
into a 100 ml Erlen-
meyer flask (NS 29/32).
3.
30 ml Sulfuric Acid
,
2.0 ml Potassium Iodide solution
and
0.3 ml Tin(II)chloride
are added to the water sample.
4. Close the flask and shake, wait for a period of
15
minutes
.
5. Prepare
2.0 g Zinc
.
6. Fill the absorption tube with exact
5.0 ml of absorption
solution
(see picture
and
; use pipette).
7. After end of the 15 minutes reaction time add the
2 g Zinc to the Erlenmeyer flask and
immediately as-
semble
the apparatus with the prepared absorption
tube (see picture
).
8. The reaction starts
(fume hood!)
.
Wait for
60 minutes
reaction time.
Performing test procedure:
9. Fill a clean
20 mm cell
(note 1) with
deionized water
.
10 Place the cell in the sample chamber making sure that
the positioning is correct.
11. Press
ZERO
key.
12. After zeroing remove the cell from the sample chamber.
Empty the cell and dry completely.
13. Fill the cell with the colored absorption solution. (see
picture
).
14. Place the cell in the sample chamber making sure that
the marks are aligned.
15. Press
TEST
key.
The result is shown in the display in mg/l Arsenic.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...