250
SP600 Spectrophotometer 01/2008
1.1 Methods
Surfactants, anionic
with MERCK Spectroquant
®
Cell Test, No. 1.14697.0001
0.05 – 2 mg/l MBAS
1. Place the supplied blank in the sample chamber making
sure that the marks are aligned.
2. Press
ZERO
key.
3. Remove the vial from the sample chamber.
4. Pipette
5 ml of water sample
into one reaction
tube.
Do not mix contents!
5. Fill the same reaction tube with the same size drops by
holding the bottle vertically and squeeze slowly:
3 drops reagent T-1K
Do not mix contents!
6. Add
3 drops reagent T-2K
.
7. Close the vial tightly with the cap and
shake for 30
seconds
.
8. Press
[ ]
key.
Wait for a
reaction period of 10 minutes
.
After reaction period is finished proceed as follows:
Swirl the vial
and than place the vial in the sample
chamber making sure that the marks are aligned.
9. Press
TEST
key.
The result is shown in the display in mg/l MBAS.
Ø 16 mm
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
start:
3
7
5
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...