SP600 Spectrophotometer 01/2008
Store user-calibration
Perform the required method as described in the manual
using a standard of known concentration instead of the
water sample.
If the test result is displayed press [MODE] [4] [5] keys and
confirm with
[ ]
key.
The display shows:
Pressing the arrow key [ ] once increases the displayed
result.
Pressing the arrow key [ ] once decreases the displayed
result.
Press keys till the displayed result corresponds to the value
of the standard.
Confirm with
[ ]
key to store the new calibration factor.
Cancel user calibration by pressing [ESC] key.
The display shows:
Now the method name is displayed inverse and the test
result is calculated with the new calibration factor.
100 Chlorine T
0.02-6 mg/l Cl2
0.90 mg/l free Cl2
Jus Factor
saved
100
Chlorine T
0.02-6 mg/l Cl2
1.00 mg/l free Cl2
<user calibration>
100 Chlorine T
0.02-6 mg/l Cl2
0.90 mg/l free Cl2
up: , down:
save:
Mode
4
5
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...