139
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing
substances, the subsequent determination of Hydrogen peroxide may show lower
results. To avoid any measurement errors, only use glassware free of Chlorine
consumption.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2. Preparing the sample:
When preparing the sample, the escape of Hydrogen peroxide gases, e.g. by pipetting
or shaking, must be avoided. The analysis must take place immediately after taking
the sample.
3. The DPD colour development is carried out at a pH value of 6.3 to 6.5. The reagent
tablet therefore contains a buffer for the pH adjustment.
Strong alkaline or acid water samples must be adjusted between pH 6 and pH 7 before
the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding of the measuring range:
Concentrations above 5 mg/l Hydrogen peroxide can produce results within the
measuring range up to 0 mg/l. In this event, the water sample must be diluted with
water free of Hydrogen peroxide. 10 ml of the diluted sample will be mixed with the
reagent and the measurement repeated.
Oxidizing agents such as Chlorine, Ozone etc. interfere as they react like Hydrogen
peroxide.
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...