199
SP600 Spectrophotometer 01/2008
1.1 Methods
15.
Transfer the contents of the second beaker into
the prepared beaker.
Dissolve tablets.
16. Fill the 50 mm cell with the colored test solution.
17. Place the cell in the sample chamber making sure that
the positioning is correct.
18. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement starts
automatically.
The result is shown in the display in:
mg/l Ozone
mg/l total Chlorine
Notes:
see page 197
*,** mg/l O
3
*,** mg/l tot Cl
T1 accepted
prepare T2
press TEST
Countdown
2:00