200
SP600 Spectrophotometer 01/2008
1.1 Methods
Ozone, in absence of Chlorine
with Tablet
0.02 – 0.5 mg/l O
3
1. Fill a clean 50 mm cell with
water sample
.
2. Place the cell in the sample chamber making sure that
the positioning is correct.
3. Press
ZERO
key.
4. Remove the cell from the sample chamber. Empty the
cell and dry completely.
5. Rinse a beaker with the water sample and
empty it,
leaving a few drops remaining in the beaker
.
6. Add
one DPD No. 1 tablet
and
one DPD No. 3 tablet
straight from the foil and crush the tablet using a clean
stirring rod.
7. Add
10 ml of water sample
and dissolve the tablets.
8. Fill the 50 mm cell with the colored test solution.
9. Place the cell in the sample chamber making sure that
the positioning is correct.
10. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Ozone.
Notes:
see page 197
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
2:00
2
9
9
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...