165
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. This test is applicable for the determination of soluble Manganese in water and
wastewater.
2. Highly buffered or water sample with extreme pH-values may exceed the buffering
capacity of the reagents and requires sample pre-treatment.
If samples were acidified for storing, adjust the pH between 4 and 5 with 5 mol/l (5 N)
Sodium hydroxide before test. Do not exceed pH 5, as manganese may precipitate.
3. Interferences:
Interfering substance
Interference level
Calcium
greater than 700 mg/l
Chloride
greater than 70 000 mg/l
Iron
greater than 5 mg/l
Magnesium
greater than 100 000 mg/l
4. Mn
MnO
4
KMnO
4