131
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. This colour scale was originally developed by A. Hazen as a visual comparison scale.
It is therefore necessary to ascertain whether the extinction maximum of the water
sample is in the range from 420 to 470 nm, as this method is only suitable for water
samples with yellowish to yellowish-brown coloration. Where applicable, a decision
should be made based on visual inspection of the water sample.
2. This method 203 – Hazen – is calibrated on the basis of the standards specified by
”Standard Methods for the Examination of Water and Wastewater” (also see EN ISO
7887:1994).
1 Pt-Co colour unit = 1 mg/L of platinum as chloroplatinate ion
3. Sample taking, preservation and storage:
Pour the water sample into clean glass or plastic containers and analyse as soon as
possible after the sample is taken. If this is not possible, fill the container right up to the
top and seal tightly. Do not stir the sample; avoid lengthy contact with the air.
The sample may be stored in a dark place at a temperature of 4°C for 24 hours. Before
performing measurements, the water sample must be brought up to room temperature.
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...