176
SP600 Spectrophotometer 01/2008
1.1 Methods
Nitrate LR
with Tube Test
0.5 – 14 mg/l N
1. Place the supplied blank (red label) in the sample
chamber making sure that the marks are aligned.
2. Press
ZERO
key.
3. Remove the vial from the sample chamber.
4. Add
0.5 ml of water sample
into one reaction tube.
5. Close the vial tightly with the cap and invert several
times to mix the contents.
(Caution: tube becomes warm!)
6. Add
0.2 ml Nitrate-111
.
7. Close the vial tightly with the cap and invert several
times to mix the contents.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press
TEST
key.
Wait for a
reaction period of 15 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in mg/l Nitrate.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
Countdown
15:00
2
6
7
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...