238
SP600 Spectrophotometer 01/2008
1.1 Methods
Silica LR / Silicon dioxide LR
with Powder Pack
and Liquid Reagent
0.1 – 1.6 mg/l SiO
2
Use two clean vials (24 mm Ø) and mark one as blank for
zeroing.
1. Fill each vial with
10 ml of water sample
.
2. Add
0.5 ml Vario Molybdate 3 reagent solution
into
each vial.
3. Close the vials tightly and swirl several times to mix the
contents (Note 1).
4. Press
[ ]
key.
Wait for a
reaction period of 4 minutes
(Note 2).
After reaction period is finished proceed as follows:
5. Add
one Vario Silica Citric Acid F10 Powder Pack
straight from the foil into each vial.
6. Close the vials tightly and swirl several times to mix the
contents.
7. Press
[ ]
key.
Wait for a
reaction period of 1 minute
(Note 3).
After reaction period is finished proceed as follows:
8. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
9. Add
one Vario LR Silica Amino Acid F F10 Powder
Pack
straight from the foil into the vial (the sample).
10. Close the vial tightly with the cap and swirl several
times to mix the contents.
Countdown
4:00
start:
Countdown
1:00
start:
1
2
3
4
3
5
1
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...