1.1 Methods
Spectral Absorption Coefficient
(S Abs)
0 – 50 m
-1
Spectral Absorption Coefficient
at 436 nm (S Abs1)
Spectral Absorption Coefficient
at 525 nm (S Abs2)
Spectral Absorption Coefficient
at 620 nm (S Abs3)
Methods 345, 346 and 347 are called up one after the
other and the water sample is analysed using the tests as
described below:
Sample preparation:
1. Filter the water sample through a membrane filter with
a pore width of 0.45 μm. (At least 100 ml of the water
sample should be filtered.)
Performing test procedure:
2. Fill a clean 50 mm cell with
deionized water
(Note 1).
3. Place the cell in the sample chamber making sure that
the positioning is correct.
4. Press
ZERO
key.
5. Remove the cell from the sample chamber and empty
the cell.
6. Use some of the filtered water sample to rinse out the
cell, then fill the sample into the cell.
7. Immediately place the cell in the sample chamber ma-
king sure that the positioning is correct.
8. Press
TEST
key.
The result is shown in the display in (m
-1
).
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
3
4
6
3
4
5
3
4
7
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...