314
SP600 Spectrophotometer 01/2008
2.4.8 Special functions
Langelier Saturation Index (Water Balance)
For calculation the following tests are required:
• pH-value
• Temperature
• Calcium hardness
• Total Alkalinity
• TDS (Total Dissolved Solids)
Run the test separately and note the results.
Calculate the Langelier Saturation Index as described:
Calculation of Langelier Saturation Index
With Mode 71 (see below) it is possible to select between
degree Celsius or degree Fahrenheit.
Press [MODE] [7] [0] keys.
Confirm with
[ ]
key.
The display shows:
Enter the temperature value (T) in the range
between 3 and 53°C and confirm with
[ ]
key.
If °F was selected, enter the temperature value in the range
between 37 und 128°F.
The display shows:
Enter the value for Calcium hardness (CH) in the range
between 50 and 1000 mg/l CaCO
3
and confirm with
[ ]
key.
The display shows:
Enter the value for Total Alkalinity (TA) in the range between
5 and 800 mg/l CaCO
3
and confirm with
[ ]
key.
The display shows:
Enter the value for TDS (Total Dissolved Solids) in the range
between 0 und 6000 mg/l and confirm with
[ ]
key.
calcium hardness
50<=CH<=1000
+_ _ _ _
<Langelier>
temperature °C:
3°C <=T<=53°C
+_ _ _ _
tot. alkalinity
5<=TA<=800
+_ _ _ _
total dissol. solids
0<=TDS<=6000
+_ _ _ _
Mode
7
0
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...