119
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. The same batch of SPADNS reagent solution must be used for adjustment and test.
The adjustment process needs to be performed for each new batch of SPANDS reagent
solution (see Standard methods 20th, 1998, APHA, AWWA, WEF 4500 F D., S. 4-82).
The procedure is described in chapter 2.4.5 "Calibration" Mode 40 on page 291.
2. During adjustment and test the same vial should be used for zeroing and test, as
different vials may exhibit minor tolerances.
3. The calibration solution and the water samples to be tested should have the same
temperature (± 1°C).
4. As the test result is highly dependent on exact sample and reagent volumes, the sample
and reagent volumes should always be metered by using a 10 ml resp. 2 ml volumetric
pipette (class A).
5. The accuracy of the test methods decreases above a level of 1.2 mg/l Fluoride. Although
the results are sufficiently accurate for most applications, even more exact results can be
achieved by 1:1 dilution of the sample prior to use and subsequent multiplication of the
result by 2.
6. SPADNS reagent solution contains Arsenite.
Chlorine concentrations up to 5 mg/l do not interfere.
7. Seawater and wastewater samples must be distilled.
8. It is convenient to use special vials with larger volume.
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...