SP600 Spectrophotometer 01/2008
2.3.5 Performing Tests
When zero calibration is complete, remove the vial from the sample chamber
and perform the tests as described under “Method”.
When the results have been displayed:
- at some methods you can change between different chemical species
- you can store and/or print out the results
- perform further analysis with the same zero
- select a new method
2.3.6 Ensuring reaction periods (countdown)
For the compliance with reaction periods there is incorporated a time delay,
the countdown.
There are two kinds of countdowns:
•
• Press
[ ]
key.
Prepare water sample, start countdown with
[ ]
key and
proceed as described in the mode description.
The vial must not be placed in the sample chamber.
• Press
TEST
key.
Prepare the water sample as described in the method
description and place the vial in the sample chamber.
The display shows the countdown by pressing the
TEST
key and the countdown is started automatically.
The reaction period is finished the measurement starts
automatically.
Notes:
1. It is possible to finish the working countdown by pressing the
[ ]
key. Reading starts
immediately. In this case the operator is responsible for ensuring the necessary reaction
period by himself.
Non-compliance with reaction periods lead to incorrect test results.
2. The time remaining is displayed continuously.
The beeper indicates the last 10 seconds.
Countdown
1:59
Countdown
2:00
Start:
Test
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...