BeneVision N1 Patient Monitor Operator’s Manual
14 - 1
14
Monitoring Invasive Blood Pressure (IBP)
14.1
IBP Introduction
This patient monitor can monitor up to four invasive blood pressures.
IBP monitoring is intended for adult, pediatric, and neonatal patients. PAWP monitoring is available only for the
external display. PAWP monitoring is only intended for adult and pediatric patients.
WARNING
•
The IBP module cannot be used for transport outside the hospital environment.
14.2
IBP Safety Information
WARNING
•
Use only pressure transducers specified in this manual. Never reuse disposable pressure
transducers.
•
Make sure that the applied parts never contact other conductive parts.
•
To reduce the hazard of burns during high-frequency surgical procedure, ensure that the monitor’s
cables and transducers never come into contact with the high-frequency surgical units.
•
When using accessories, their operating temperature should be taken into consideration. For more
information, see instructions for use of accessories.
•
All invasive procedures involve risks to the patient. Use aseptic technique. Follow catheter
manufacturer's instructions.
•
Mechanical shock to the invasive blood pressure transducer may cause severe shifts in zero balance
and calibration, and cause erroneous readings.
(1) IBP cable connector
(2) IBP dialog hard key
(3) Zero IBP hard key
(1)
(1)
(3)
(2)
(1)