BeneVision N1 Patient Monitor Operator’s Manual
Note: When arrhythmia alarms occur, check the patient’s condition and the ECG connections.
D.1.3
Resp Physiological Alarm Messages
D.1.4
SpO
2
Physiological Alarm Messages
D.1.5
PR Physiological Alarm Messages
D.1.6
NIBP Physiological Alarm Messages
D.1.7
IBP Physiological Alarm Messages
Pacer Not Capture
Prompt
Missed Beats
Prompt
A-Fib
Prompt
Alarm message
Default priority
Alarm message
Default priority
Cause and solution
Resp Aritifact
High
The patient’s heartbeat has interfered with his respiration. Check the
patient’s condition and the Resp connections.
Zero RR
High
The respiration signal was so weak that the monitor cannot perform
respiration analysis. Check the patient’s condition, module and patient
connections.
Alarm message
Default priority
Cause and solution
SpO
2
Desat
High
The SpO
2
value falls below the desaturation alarm limit. Check the
patient’s condition and check if the alarm limit settings are correct.
Alarm message
Default priority
Cause and solution
No Pulse
High
The pulse signal was so weak that the monitor cannot perform pulse analysis.
Check the patient’s condition, SpO2 sensor and measurement site.
Alarm message
Default priority
Cause and solution
NIBP-S/NIBP-D/NIBP-
M Extremely High
High
The NIBP value is higher than the NIBP Extreme alarm high limit. Check the
patient’s condition and check if the alarm limit settings are correct.
NIBP-S/NIBP-D/NIBP-
M Extremely Low
High
The NIBP value is lower than the NIBP Extreme alarm low limit. Check the
patient’s condition and check if the alarm limit settings are correct.
Alarm message
Default priority
Cause and solution
Art-S/Art-D/Art-M
Extremely High
High
The Art value is higher than the Art Extreme alarm high limit. Check the
patient’s condition and check if the alarm limit settings are correct.
Art-S/Art-D/Art-M
Extremely Low
High
The Art value is lower than the Art Extreme alarm low limit. Check the patient’s
condition and check if the alarm limit settings are correct.