BeneVision N1 Patient Monitor Operator’s Manual
C - 3
C.1.2.2
Arrhythmia Threshold Default Settings
C.1.3
ST Default Settings
Item
Default Setting
Adult
Pediatric
Neonate
Asystole Delay
5 sec
5 sec
5 sec
Tachy
120 bpm
160 bpm
200 bpm
Brady
50 bpm
75 bpm
100 bpm
Extreme Tachy
160 bpm
180 bpm
220 bpm
Extreme Brady
35 bpm
50 bpm
60 bpm
Multiform PVC Window
15 beats
15 beats
15 beats
PVCs/min
10
10
10
Pauses/min
8
8
8
Pause Threshold
2.0 sec
2.0 sec
2.0 sec
AF/Irr Rhy End Time
2 min
2 min
2 min
V-Tach Rate
130 bpm
130 bpm
160 bpm
V Brady Rate
40 bpm
40 bpm
40 bpm
V-Tach PVCs
6
6
6
V-Brady PVCs
5
5
5
Item
Default Setting
ST Alarm Mode
Absolute
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb (ST
Alarm Mode set to
Absolute
)
Alarm switch
Off
High limit
0.2 mV
Low limit
-0.2 mV
Priority
Med
Alarm Outputs
Off
ST Single,
ST Dual (ST Alarm Mode set to
Relative
)
Alarm switch
Off
High limit
0.1 mV
Low limit
-0.1 mV
Priority
Med
Alarm Outputs
Off
ST Analysis
Off
ST Segment
Auto
Show Marker
Off
ST Point
J+60
Auto Adjust
On
J
48
ISO
-80