BeneVision N1 Patient Monitor Operator’s Manual
17 - 9
17.3.3.2
Marking Implemented Items
Check off implemented items. Then the time and date are automatically recorded and displayed.
■
You can select the
symbol to change the date and time.
■
Select
Reset
to clear the current results.
17.3.4
Reviewing SepsisSight Trend Data
Select the
Graphic Trends
tab to view the trend of parameters of resuscitation.
When a recommended treatment is checked off on the
Part I
page and
Part II
page, relevant event is marked in
the tabular trend. Vertical lines of different colors indicate the event type:
■
White: inspection performed
■
Blue: medication
■
Green: goal achieved
■
Purple: other treatment
17.4
Rescuing Mode
You can put the monitor into the rescue mode when rescuing a patient. In the rescue mode, the monitor displays
the following information:
■
Values and waveforms of physiological parameters
■
CPR
Dashboard
You can output the rescue report.
The rescue mode is intended for adult and pediatric patients
.
WARNING
•
The rescue mode is not intended for neonatal patients.
•
In the rescue mode, all physiological alarms and part of technical alarms are disabled.
NOTE
•
Licenses are required for the CPR Dashboard function.
•
A responsible nurse is required to record the rescue process. Recording shall not affect patient
rescue.
17.4.1
Entering the Rescue Mode
To enter the rescue mode, choose either of the following ways
:
■
Select the
Rescue Mode
quick key.
■
Select the
Main Menu
quick key
→
from the
Alarm
column select
Rescue Mode
.
17.4.2
CPR Dashboard
The CPR Dashboard helps you record the medications and treatments administrated during patient rescue. You
can record the following information on the monitor:
■
Rescue start time and end time
■
The use of drugs, for example adenaline, amiodarone and other drugs
■
The administrated treatments, for example, CPR, defibrillation, and other treatments