BeneVision N1 Patient Monitor Operator’s Manual
D - 9
Loading Default Config
Failed
Low
A
The default configuration is not correctly loaded.
The monitor will restore to the factory default
configuration for the current patient category.
Read dock E2PROM error!
High
C
1. Check if you’re using the specified external
display.
•
If you’re using the specified external display,
remove the N1 from the Dock, and reconnect
the N1 and the Dock.
•
If you’re not using the specified external
display, replace current external display with
the specified external display. Then remove
the N1 from the Dock, and reconnect the N1
and the Dock.
2. If the alarm persists, contact your service
personnel.
XX Conflicts
(XX refers to the module
label)
Prompt
/
The same type of corresponding module being
used exceeds the supported number. Remove
the conflict module.
XXX Measurement has
been closed
(XX refers to the module
label)
Prompt
/
The parameter module is disabled. Switch on the
module if you want to use it. For more
information, see
.
The display setup for XXX
is disabled.
(XX refers to the
parameter label)
Prompt
/
The parameter of the newly inserted module is
not displayed on the screen. Select a desired area
to display the parameter numerics and
waveforms. For more information, see
3.11.2 Displaying Parameter Numerics and
Waveforms
.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution