BeneVision N1 Patient Monitor Operator’s Manual
3 - 11
3.7.3.1
Entering the Privacy Mode
To enter the privacy mode, select the
Main Menu
quick key
→
from the
Display
column select
Privacy Mode
→
select
Ok
.
The monitor has the following features after entering the privacy mode:
■
The screen turns blank.
■
Except for the low battery alarm, the monitor inactivate alarm tone and alarm light of all other alarms.
■
The monitor suppresses all system sounds, including heart beat tone, pulse tone, and prompt tone.
WARNING
.
•
In Privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor.
Alarms are presented only at the CMS. Pay attention to potential risk.
NOTE
•
The privacy mode is not available if the Department is set to OR.
•
You cannot enter the privacy mode if a low battery alarm occurs.
3.7.3.2
Exiting the Privacy Mode
The monitor automatically exit the privacy mode in any of the following situations:
■
The monitor disconnects from the CMS.
■
The low battery alarm occurs.
You can also operate the touchscreen, mouse, or keyboard to manually exit the privacy mode.
3.7.4
Night Mode
The night mode is a special clinical monitoring mode. To avoid disturbing the patient, you can use the night
mode.
3.7.4.1
Entering the Night Mode
To enter the night mode, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Display
column select
Night Mode
.
2.
Change the night mode settings if necessary.
3.
Select
Enter Night Mode
.
The night mode settings are as follows by default:
■
Brightness: 1
■
Alarm Volume: 2
■
QRS Volume: 1
■
Key Volume: 0
■
NIBP End Tone: Off
■
Stop NIBP: Off
CAUTION
•
Verify the night mode settings before entering the night mode. Pay attention to the potential risk if
the setting value is low.
3.7.4.2
Exiting the Night Mode
To cancel the night mode, follow this procedure: