7 - 12
BeneVision N1 Patient Monitor Operator’s Manual
7.8.1
Resetting Physiological Alarms
Physiological alarms give different alarm indicators when the alarm system is reset:
■
The alarm sound is silenced.
■
A √ appears before the alarm message, indicating that the alarm is acknowledged.
■
The color of the parameter numeric background corresponds with the alarm priority, but the parameter
numeric does not flash.
7.8.2
Resetting Technical Alarms
Technical alarms give different alarm indicators when the alarm system is reset:
■
Some technical alarms are cleared. The monitor gives no alarm indications.
■
Some technical alarms are changed to the prompt messages.
■
For some technical alarms, the alarm is silenced and a √ appears before the alarm message, indicating that
the alarm is acknowledged.
For details about the indications of technical alarms when the alarm system is reset, see
7.8.3
Setting Alarm Light Status on Alarm Reset
When the alarm system is reset, the monitor presents the alarm light by default , but you can switch off the alarm
light. To change the alarm light setting, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Pause/Reset
tab.
3.
Set
Alarm Light
.
◆
On When Reset
: when the alarm system is reset, the alarm tones of the current alarms are switched
off, but the alarm lamp remains flashing.
◆
Off When Reset
: when the alarm system is reset, both the alarm tone and alarm lamp of the current
alarms are switched off.
7.9
Latching Alarms
The latching setting for physiological alarms defines how alarm indicators behave if you do not reset the alarms.
■
If you do not “latch” physiological alarms, their alarm indications disappear when the alarm condition ends.
■
If you “latch” physiological alarms, all visual and audible alarm indications remains until you reset the
alarms. For latched alarms the time when the alarm is last triggered is displayed behind the alarm message.
You can separately latch visual indications or simultaneously latch the visual and the audible indications.
■
When visual indications are latched, visual indications, including alarm lamp, alarm message and its
background remains when the alarm condition ends and the time when the alarm last triggered is
displayed behind the alarm message.
■
When audible indications are latched, the monitor continues to issues alarm sounds when the alarm
condition ends.
To latch physiological alarms, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Latching
tab.
3.
Select how you want to latch the alarms.
Alarm latching rules are as follows:
■
You can separately latch visual alarm signal.
■
Latching audible alarm signal simultaneously latches visual signal.
■
Selecting alarms of lower priority simultaneously latches higher priority alarms.