14 - 8
BeneVision N1 Patient Monitor Operator’s Manual
14.7.1
PAWP Equipment to Patient Connection
14.7.2
Preparing to Measure PAWP
To prepare to monitor PAWP, follow this procedure:
1.
Connect the IBP transducer, the IBP cable and the monitor. For more information, see
.
2.
Follow the manufacturer’s instructions to connect the PA port of the thermodilution catheter and the
patient end of the IBP transducer.
3.
Zero the IBP transducer. For more information, see
14.3.3 Zeroing the IBP transducer
.
4.
Set the IBP label to
PA
since the PAWP is measured on PA. For more information, see
.
14.7.3
Measuring PAWP
To measure the PAWP, follow this procedure:
1.
Select the PA numeric area or waveform area to enter the
PA
menu, and then select
PAWP
.
2.
Wedge the flotation catheter into the pulmonary artery by observing the PA waveform changes on the
screen, referring to the following figure.
(1) N1 monitor
(2) Flush bag
(3) IBP transducer
(4) Three-way valve
(5) PA distal port
(6) Balloon inflation valve
(7) Thermodilution catheter
(8) Balloon
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)