13 - 4
BeneVision N1 Patient Monitor Operator’s Manual
4.
Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass
unrestricted through the tubing.
CAUTION
•
A wrong cuff size and a folded or twisted bladder can cause inaccurate measurements.
•
Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.
This may cause inaccurate blood pressure values.
•
Use care when placing the cuff on an extremity used for monitoring other patient parameters.
13.7
Starting and Stopping NIBP Measurements
Start and stop NIBP measurement by selecting the NIBP quick keys or from the NIBP menu.
13.8
Changing NIBP Settings
13.8.1
Setting the NIBP Alarm Properties
To set the NIBP alarm properties, follow this procedure:
1.
Select the NIBP numeric area to enter the
NIBP
menu.
2.
Select the
Alarm
tab.
3.
Enter the password if required. For more information, refer to
6.9.3 Selecting Password for User
.
4.
Set alarm properties as desired.
13.8.2
Setting the Initial Cuff Inflation Pressure
To set initial cuff inflation pressure, follow this procedure:
1.
Select the NIBP numeric area to enter the
NIBP
menu.
Task
By Quick Key
From NIBP menu
Start a manual measurement
NIBP Start/Stop
quick key
Start NIBP
button
Start auto NIBP series
NIBP Start/Stop
quick key
Make sure to set the
Interval
before
starting the auto NIBP.
Setup
tab
→
set
Interval
→
Start
NIBP
button
NIBP Measure
quick key
→
select
interval
Start
NIBP sequence measurement
NIBP Measure
quick key
→
Sequence
Sequence
tab
→
set NIBP sequence
→
Start NIBP
button
Start STAT measurement
NIBP Measure
quick key
→
STAT
STAT
button
Stop the current NIBP
measurements
NIBP Start/Stop
quick key
Stop NIBP
button
End auto NIBP series or NIBP
Sequence
/
Stop All
button
Stop STAT measurement and end
series
NIBP Start/Stop
quick key
Stop NIBP
or
Stop All
button