C - 2
BeneVision N1 Patient Monitor Operator’s Manual
C.1.2
Arrhythmia Default Settings
C.1.2.1
Arrhythmia Alarm Default Settings
Item
Alarm Switch
Priority
Alarm Outputs
Asystole
On
High, unadjustable
Off
V-Fib/V-Tach
On
High, unadjustable
Off
V-Tach
On
High, unadjustable
Off
Vent Brady
On
High, unadjustable
Off
Extreme Tachy
On
High, unadjustable
Off
Extreme Brady
On
High, unadjustable
Off
R on T
CCU: On
Other departments: Off
Med
Off
Run PVCs
Off
Low
Off
Couplet
Off
Prompt
Off
Multiform PVC
Off
Med
Off
PVC
Off
Prompt
Off
Bigeminy
CCU: On
Other departments: Off
Med
Off
Trigeminy
CCU: On
Other departments: Off
Med
Off
Tachy
Off
Med
Off
Brady
Off
Med
Off
Pacer Not Capture
Off
Promt
Off
Pacer Not Pacing
Off
Promt
Off
Missed Beats
Off
Promt
Off
Nonsus V-Tach
CCU: On
Other departments: Off
Med
Off
Vent Rhythm
CCU: On
Other departments: Off
Med
Off
Pause
Off
Low
Off
Irr. Rhythm
Off
Promt
Off
A-Fib
Off
Promt
Off
PVCs/min
CCU: On
Other departments: Off
Med
Off
Pauses/min
CCU: On
Other departments: Off
Med
Off