BeneVision N1 Patient Monitor Operator’s Manual
9 - 3
9.5.2
Setting Tachycardia and Bradycardia Thresholds (For Glasgow Algorithms)
To set tachycardia and bradycardia thresholds, follow this procedure:
1.
On the ECG 12-Lead screen, select
Setup
to enter the
12-Lead Setup
menu.
2.
Select the
Setup
tab.
3.
Set
Tachy
and
Brady
.
NOTE
•
The tachycardia threshold only applies to patients whose age exceeds 180 days.
•
The bradycardia threshold only applies to patients whose age exceeds 2191 days.
9.5.3
Setting the 12-Lead Interpretation Report (For Glasgow Algorithms)
To set the 12-lead interpretation report, follow this procedure:
1.
On the ECG 12-Lead screen, select
Setup
to enter the
12-Lead Setup
menu.
2.
Select the
Report
tab.
3.
Set the format and items included in the 12-lead interpretation report.
9.6
Saving the 12-Lead Interpretation Report
At the completion of 12-lead ECG interpretation, select
Save
to save the report. You can review the saved 12-
lead interpretation reports. For more information, see
16.2.10 12-Lead ECG Review Page
9.7
Printing the 12-Lead Interpretation Report
At the completion of 12-lead ECG interpretation, select
to output the report via the printer.
9.8
Exiting the ECG 12-Lead Screen
To exit the ECG 12-Lead screen, select
Exit
on the ECG 12-Lead screen.