BeneVision N1 Patient Monitor Operator’s Manual
D - 1
D
Alarm Messages
D.1
Physiological Alarm Messages
This section lists major physiological alarms, their default priority, and the actions that can be taken when an
alarm occurs. Some of alarms may not be included.
D.1.1
General Physiological Alarm Messages
Note: XX represents a measurement or parameter label, such as HR, NIBP, PVCs, RR, SpO
2
, PR, and so on.
D.1.2
Arrhythmia Alarm Messages
Alarm messages
Default priority
Cause and solution
XX High
High/Med, configurable
XX value has risen above the high alarm limit or fallen below the low
alarm limit. Check the patient’s condition and check if the patient
category and alarm limit settings are correct.
XX Low
High/Med, configurable
Alarm message
Default priority
Asystole
High
V-Fib/V-Tach
High
V-Tach
High
Vent Brady
High
Extreme Tachy
High
Extreme Brady
High
PVCs/min
Med
Pauses/min
Med
R on T
Med
Bigeminy
Med
Trigeminy
Med
Tachy
Med
Brady
Med
Multiform PVC
Med
Vent Rhythm
Med
Nonsus V-Tach
Med
Run PVCs
Low
Pause
Low
Couplet
Prompt
PVC
Prompt
Irr Rhythm
Prompt
Pacer Not Pacing
Prompt