BeneVision N1 Patient Monitor Operator’s Manual
7 - 7
7.6.4
Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
To set the alarm delay time, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Other
tab.
3.
Set
Alarm Delay
.
The setting of
Alarm Delay
is not applied to the zero RR alarms and the ST alarms. You can set
Zero RR Delay
and
ST Alarm Delay
separately.
7.6.4.1
Setting the Zero RR Delay Time
To set the Zero RR delay time, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Select
Zero RR Delay
to set the zero RR delay time.
7.6.4.2
Setting the
ST Alarm Delay Time
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select
.
2.
Select the
Alarm
tab
→
Other
tab.
3.
Set
ST Alarm Delay
.
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4
(Venous
pressure)
IBP-M
Mean × 0.75
Mean × 0.75
Mean × 1.25
Mean × 1.25
3 to 40
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29
33 to 35: 29
33 to 35: 41
33 to 35: 41
Same as the
measurement range
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
+ 6)
36 to 45:
(etCO
2
+ 6)
Same as the
measurement range
46 to 48: 39
46 to 48: 39
46 to 48: 51
46 to 48: 51
Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None
None
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
Module
Parameter
Lower Limit
Upper Limit
Auto Limit Range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate