BeneVision N1 Patient Monitor Operator’s Manual
D - 5
D.2.5
Temp Technical Alarm Messages
D.2.6
NIBP Technical Alarm Messages
D.2.7
IBP Technical Alarm Messages
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
T1/T2 Sensor Off
Low
A
Check the sensor connection and reconnect the
sensor.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
NIBP Cuff Loose
Low
A
There is a leak in the cuff or air tubing. Use a cuff
of correct type based on the patient size. Apply
the cuff and connect the air tubing as instructed
in the manual.
NIBP Cuff or Airway Leak
Low
A
Check the NIBP cuff and pump for leakages.
NIBP Airway Error
Low
A
The air tubing may be occluded. Check the air
tubing for an occlusion or kinking. If the alarm
persists, contact your service personnel.
NIBP Weak Signal
Low
A
The patient’s pulse is weak or the cuff is loose.
Check the patient’s condition and replace the cuff
application site.
NIBP Overrange
Low
A
The measured NIBP value exceeds the module
measurement range. Check the patient’s
condition.
NIBP Excessive Motion
Low
A
Check the patient’s condition and reduce patient
motion.
NIBP Cuff Overpressure
Low
A
The NIBP airway may be occluded. Check the
airway and measure again. If the alarm persists,
contact your service personnel.
NIBP Timeout
Low
A
The measurement time exceeds 120 seconds in
the adult or pediatric mode, or exceeds 90
seconds in the neonatal mode, and the BP value
cannot be obtained. Check the patient’s
condition and NIBP connections, or replace the
cuff and measure again.
NIBP Cuff and Patient
Mismatch
Low
A
The cuff type mismatches the patient category.
Verify the patient category or replace the cuff if
necessary. If patient catergory is correct, check
that the tubing is not bent and the airway is not
occluded.
NIBP Airway Leak
Low
A
Airway leakage is found during the NIBP leakage
test. Check the NIBP cuff and pump for leakages.
Alarm message
Default priority
Indication on
alarm reset
Cause and solution
XX Sensor Error
Med
C
The IBP sensor fails. Replace the sensor.
XX No Sensor
High, Med, or
Low,
configurable
A
The IBP patient cable and/or corresponding IBP
sensor is not connected or detached. Check the
cable and sensor connection.