BeneVision N1 Patient Monitor Operator’s Manual
C - 5
C.3
SpO
2
Default Settings
Zero RR
Alarm switch
On
Priority
High, unadjustable
Alarm Outputs
Off
Zero RR Delay
Adult: 20 sec
Pediatric: 20 sec
Neonate: 15 sec
RR Source
Auto
Resp Lead
Adult: Auto
Pediatric: Auto
Neonate: II
Gain
×2
Speed
6.25 mm/s
Auto Threshold Detection
On
Item
Default Setting
Item
Default Setting
SpO
2
Alarm
switch
On
High limit
Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit
90%
Priority
Med
Alarm Outputs
Off
SpO
2
Desat
Alarm switch
On
Low limit
80%
Priority
High
Alarm Outputs
Off
Satsecond (for Nellcor SpO
2
)
Off
NIBP Simul
Off
Sensitivity (for Mindray SpO
2
)
Medium
Display PI (for Mindray SpO
2
and Masimo
SpO
2
)
On
Speed
25 mm/s