BeneVision N1 Patient Monitor Operator’s Manual
3.13.3
External Display Troubleshooting
CAUTION
•
Use only specified display. Using unspecified display may result in unknown problem.
3.14
The OxyCRG Screen
(only available for the independent external display)
The OxyCRG screen is the default user screen when the neonatology department is selected. It displays 6-minute
HR/btbHR, SpO
2
trends, CO
2
/Resp compressed waveform, ABD parameters, and the latest ABD events.
The OxyCRG function is intended for neonatal patents only.
3.14.1
Entering the OxyCRG Screen
To enter the OxyCRG screen, choose any of the following ways:
■
Swipe left or right on the touchscreen with two fingers to switch to the OxyCRG screen.
■
Select the
OxyCRG
quick key.
■
Select the
Screen Setup
quick key
→
select the
Choose Screen
tab
→
select
OxyCRG
.
■
Select the
Main Menu
quick key
→
from the
Display
column select
Choose Screen
→
select
OxyCRG
.
3.14.2
OxyCRG Events
The following table lists the ABD events and their criteria:
NOTE
•
The monitor records all ABD events for OxyCRG review, but only red events displays in the ABD list of
the OxyCRG screen.
Problem
Corrective Actions
Image offset
Adjust the external display by using the auto adjust function or adjust the
external display manually.
No image or the image displays abnormally
•
Check that the external display is properly connected to the AC mains
and is powered on.
•
Check that the VGA cable is properly connected.
•
Remove the N1 from the Dock and reconnect it if the problem persists.
Touchscreen failure
Check that if both ends of the USB cable accompanying the external display
are connected properly to the Dock and the external display.
Event type
Description
Remarks
A
Zero RR event: the zero RR duration exceeds the threshold.
•
A20: the zero RR duration is greater or equal to 20 seconds.
•
A15: the zero RR duration is between 15 to 20 seconds (excluding 20 seconds).
•
A10: the zero RR duration is between 10 to 15 seconds (excluding 15 seconds).
A20 is a red event
B
Low heart rate event: the duration of low heart rate exceeds the threshold.
/
D
Low SpO2 event: the duration of SpO2 Desat exceeds the threshold.
/
BD
Low heart rate and low SpO2 happen at the same time.
/
AB
Zero RR and low heart rate happen at the same time.
Red event
AD
Low heart rate and low SpO2 happen at the same time.
Red event
ABD
Zero RR, low heart rate, and low SpO2 happen at the same time.
Red event