C - 8
BeneVision N1 Patient Monitor Operator’s Manual
C.6
IBP Default Settings
Initial Pressure
Adult: 160 mmHg
Pediatric: 140 mmHg
Neonate: 90 mmHg
Interval
OR: 5 min
NICU: 30 min
Other departments: 15 min
Start Mode
Clock
NIBP End Tone
Off
Venipuncture Pressure
Auto
Item
Default Setting
Item
Default Setting
IBP-S
Alarm switch
On
High limit
■
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
■
PA
Adult: 35 mmHg
Pediatric and neonate: 60 mmHg
Low limit
■
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 55 mmHg
■
PA
Adult: 10 mmHg
Pediatric and neonate: 24 mmHg
Priority
Med
Alarm Outputs
Off