12 - 2
BeneVision N1 Patient Monitor Operator’s Manual
12.5
Changing Temperature Settings
12.5.1
Setting the Temperature Alarm Properties
To set the temperature alarm properties, follow this procedure:
1.
Select the temperature numeric area to enter the
Temp
menu.
2.
Select the
Alarm
tab.
3.
Enter the password if required. For more information, refer to
6.9.3 Selecting Password for User
.
4.
Set the alarm properties.
12.5.2
Selecting the Temperature Label
Select the temperature label according to the measurement site. To do so, follow this procedure:
1.
Select the temperature numeric area to enter the
Temp
menu.
2.
Select the
Setup
tab.
3.
Set the temperature label.
12.5.3
Displaying the Temperature Difference
To display the temperature difference between two measurement sites monitored by the same temperature
module, switch on corresponding
∆
T. To do so, follow this procedure:
1.
Select the temperature numeric area to enter the
Temp
menu.
2.
Select the
Setup
tab.
3.
Switch on
∆
T
.
12.6
Temperature Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
NOTE
•
For the physiological and technical alarm messages, see
Problem
Solution
Do not see Temp numeric area on the main
screen
1. Check that the Temp
is set to display in the
Screen Setup
menu.
For more information, see
3.11.2 Displaying Parameter Numerics and
2. Check that if the Temp parameter switch is enabled. If not, enable
the Temp measurement. For more information, see
.
3. Check that the connections of the temperature probe and the
temperature cable are tight.
Measurement fails/’--’ is displayed in the Temp
numeric area
1. If you are using a disposable probe, check the connection between
the probe and the temperature cable.
2. Try using a known good probe in case the sensor is damaged.
The
tympanic
thermometer display is frozen.
Install or remove the probe cover to activate the thermometer.