7 - 4
BeneVision N1 Patient Monitor Operator’s Manual
7.6.2
Setting Alarm Tone Properties
7.6.2.1
Changing the Alarm Volume
To change the alarm volume, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Alarm
column select
Setup.
2.
Set
Alarm Volume
. The optional alarm volume is between X to 10, in which X is the minimum volume,
depending on the setting of minimum alarm volume, and 10 is the maximum volume.
3.
Select
High Alarm Volume
to set the volume of the high priority alarm.
4.
Select
Reminder Volume
to set the volume of the reminder tone.
NOTE
•
When the alarm volume is set to 0, the alarm sound is turned off and the audio off symbol appears
on the screen.
•
You cannot set the volume of high priority alarms if Alarm Volume is set to 0.
7.6.2.2
Setting the Minimum Alarm Volume
To set the minimum alarm volume, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Audio
tab.
3.
Set
Minimum Alarm Volume
.
NOTE
•
When your monitor is connected to the CMS and Minimum Alarm Volume is set to 0, Minimum Alarm
Volume automatically changes to 2 if the CMS is disconnected.
7.6.2.3
Setting the Audible Alarm Tone Pattern
You can distinguishes the heart beat tone, pulse tone, and keystroke tone by frequency. You can choose any of
the three tone patterns: ISO, Mode 1, and Mode 2. For more information, see
.
To set the alarm tone pattern, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Audio
tab.
3.
Set
Alarm Sound
to
ISO
,
Mode 1
or
Mode 2
.
7.6.2.4
Setting the Interval between Alarm Sounds
If you choose the ISO pattern, you can change the interval between alarm tones. To change the interval between
alarm tones, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Alarm
tab
→
Audio
tab.
3.
Set
High Alarm Interval
,
Med Alarm Interval
, and
Low Alarm Interval
.
7.6.2.5
Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm is reset or switched off, the monitor issues a periodical
reminder tone. The reminder tone is enabled by default.