BeneVision N1 Patient Monitor Operator’s Manual
3 - 19
3.14.3
The Display of the ABD Event Area
The ABD event area displays parameter values of currently active OxyCRG events and lists the latest red ABD
events.
3.14.4
Setting OxyCRG Parameters
Select parameter trends or compressed waveform to set parameters and the compressed waveform you want to
display. The selected parameters will be used for ABD event calculation.
3.14.5
Setting the Threshold of ABD Events
Select any parameter trend or the compressed waveform to perform the following setup:
■
Set the threshold of ABD events.
■
Set
Event Storage Format
:
◆
1 min+3 min
: stores data one minute before and three minutes after the event.
◆
3 min+1 min
: stores data three minutes before and one minute after the event.
◆
2 min+2 min
: stores data two minutes before and two minutes after the event.
The stored data includes the trends of the OxyCRG parameters, compressed waveform, alarm thresholds, NIBP,
and Temp measurements.
3.14.6
Editing ABD Events
To edit ABD events, follow this procedure:
1.
Select the
Mark
button to enter the
Mark
dialog box.
2.
Drag the event list upwards and downwards to select the desired event.
3.
Select the patient’s status when the event happens.
4.
Select
Save
.
3.15
Turning Off the Monitor
Before turn off the monitor, perform the following check:
1.
Ensure that the monitoring of the patient has been completed.
2.
Disconnect the cables and sensors from the patient.
3.
Make sure to save or clear the patient monitoring data as required.
To turn off the monitor, press and hold the power switch for 3 seconds.
CAUTION
•
Press and hold the power switch for no less than 10 seconds to forcibly shut down the monitor if it
could not be shut down normally. This may cause loss of patient data.
NOTE
•
Turning off the monitor does not disconnect the monitor form the AC mains. To completely
disconnect the power supply, unplug the power cord.
•
In case of a temporary power failure, if the power is restored within 30 minutes, monitoring will
resume with all active settings unchanged; if the monitor is without power for more than 30
minutes, the monitor behaves the same as it is normally turned off.