15 - 10
BeneVision N1 Patient Monitor Operator’s Manual
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Module
tab
→
Other
tab.
3.
Set
CO2 Flow Rate For Neo
.
15.9
Changing O
2
Settings (For Sidestream CO
2
Module Integrating O
2
)
15.9.1
Changing O
2
Alarm Settings
To change the O
2
alarm settings, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Alarm
tab.
3.
Set alarm properties as desired.
15.9.2
Changing the O
2
Unit
To change the O
2
unit, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required
password
→
select .
2.
Select the
Unit
tab.
3.
Set
O2 Unit
.
15.9.3
Setting the O
2
Waveform
To set the O
2
waveform, follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set
Speed
and
O2 Scale
of the O
2
waveform.
15.10
Setting the Gas Compensation
The presence of interfering gas affects the CO
2
measurement. To get the best possible measuring result, it is
needed to set the gas compensation. The configured concentration of the interfering gas should be in
accordance with its actual proportion.
For the microstream CO
2
module, gas compensations are not required.
WARNING
•
Make sure to use the appropriate compensations. Inappropriate compensations may cause
inaccurate measurement values and result in misdiagnosis.
For the sidestream CO
2
module, follow this procedure to set the gas compensation:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set the compensation according to the actual condition.
For the mainstream CO
2
module, follow this procedure to set the gas compensation:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select the
Setup
tab.
3.
Set the following compensation according to the actual condition.
■
Balance Gas