Synchronization Error
Explanation:
The system detected an unexpected situation. The procedure can be continued.
Corrective Action:
1. Ensure the clamps on the red, yellow, and blue lines are properly positioned.
2. Touch
to resume or an alternate phase pad as desired.
266
Vacuum Level Out of Range
Explanation:
The device has detected the vacuum level is outside the expected range. This can be caused if
there is a kink or occlusion in the vacuum line or A&A line or if the A&A line is clamped off when
the device is trying to relieve suction. All other device functions continue to operate normally.
Corrective Action:
1. Check for kinks or obstructions in the vacuum line, A&A line, or reservoir inlet ports.
2. Ensure the A&A line is not clamped off at any point.
3. Touch
OK
to close this message.
4. Restart suction by touching
Suction
.
If the problem persists, service is required.
267
Centrifuge Reset
Explanation:
The device had to reset the centrifuge to continue processing. The procedure can continue.
Corrective Action:
Touch
Continue
to continue.
268
Procedure Interrupted
Explanation:
Power was interrupted before the procedure was completed.
Corrective Action:
•Touch
Resume Procedure
to resume the procedure in progress.
•Touch
New
to start a new procedure.