Chapter 11, Troubleshooting
177
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
155
Effluent Check Required
Explanation:
The effluent line sensor continued to detect a high level of free hemoglobin (Hgb) after two
extended Wash phases. There may be an issue with tubing placement in the effluent line
sensor, the effluent line sensor may be dirty, or there may be an issue with the wash solution.
Corrective Action:
If the effluent is clear:
1. Touch
Empty Bowl
to empty the bowl.
2. At the end of the Empty phase, check the tubing for correct placement in the effluent line
sensor.
3. Clean the effluent line sensor with a soft, damp, lint-free gauze moistened with water only.
If the effluent is not clear:
1. Ensure the correct saline solution is connected to the disposable set.
2. Touch
Extend Wash
to proceed with another extended wash.
If the problem persists, service is required to resolve this issue.
157
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
158
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text