Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
174
Suction Failure
Explanation:
An issue has been detected with the internal suction. All other device functions continue to
operate normally.
Corrective Action:
1. Connect an external regulated suction source to the reservoir.
2. Touch
OK
to close this message.
Service is required to repair the internal suction.
176
Wash Must Be Confirmed
Explanation:
The red cells in the bowl may have been washed with less than the recommended volume of
saline solution.
Corrective Action:
Touch
Continue
to empty the bowl.
OR
Touch
Resume Wash
to enter Wash.
177
Yellow Line Valve Position Error
Explanation:
An issue has been detected with the yellow line valve position. The centrifuge must stop before
the cover will unlock and corrective action can be taken.
Corrective Action:
1. Clamp the red, yellow, and blue lines.
2. Open the valve module cover.
3. Check the tubing manifold placement.
4. Check the tubing placement in the valve module.
5. Close and latch the valve module cover.
6. Unclamp the lines.
7. Close the device cover.
8. Touch
Continue
to continue.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text