Manifold Not Properly Loaded
Explanation:
The device detected that the manifold is not properly loaded.
Corrective Action:
1. Check the manifold positioning.
2. Close and latch the manifold cover.
3. Close the device cover.
If the problem persists, service is required.
238
Suction Calibration Error
Explanation:
An issue has been detected with the calibration data for the device suction. Recalibration of the
device suction is required.
Corrective Action:
1. Connect an external regulated suction source to the reservoir.
2. Touch
OK
to close this message.
Service is required to recalibrate the internal suction.
239
Yellow Line Valve Position Error
Explanation:
An issue has been detected with the yellow line valve position. This could be the result of a
spilled fluid or debris causing the valve to stick.
Corrective Action:
Inspect the area around the yellow line valve and clean if necessary.
If the problem persists, service is required.