116
Chapter 6, General Operation: Sequestration
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
2. Touch
Continue
to continue to Cell Salvage. The device displays a
prompt:
Figure 61, Changing to Cell Salvage
3. Following the instructions on the screen, change to a Cell Salvage
processing set.
a. Disconnect the collection bag harness from the effluent tubing.
b. Cap the collection bag line to prevent drips.
c. Connect the effluent tubing to the waste bag.
d. Close the blood bag line clamps.
e. Remove the upper portion of the blood bag harness, using the twist-
lock connector, and cap the port using the attached caps.
f. Open the saline bag clamp.
g. Open the reservoir drain port clamp.
h. Ensure that the slide clamp on the blood bag harness is open.
i. Touch
Cell Salvage
to continue to Cell Salvage.
Completing the
Sequestration
Cycle
At the end of each Sequestration cycle, the device displays a prompt asking
you whether to sequester another unit from the blood bag, prepare for a Cell