176
Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
152
Fluid Detected in Centrifuge Well
Explanation:
The device detected fluid in the centrifuge well. There may have been a blood spill or an issue
with the fluid detector.
Corrective Action:
1. Check the effluent line for kinks and occlusions; ensure clamps are open.
2. Inspect the bowl and fluid sensor.
3. If a spill is observed, power down the device to clean the centrifuge per instructions in the
manual and replace the disposable.
4. If no spill is detected, once the centrifuge well has been dried, touch
Retry
.
If the problem persists, service is required.
153
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text