Chapter 12, Reference Information
213
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
Electromagnetic
Compatibility
Table 17, Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Cell Saver Elite system is intended for use in the electromagnetic environment specified below.
The customer or user of the Cell Saver Elite device should ensure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1
The Cell Saver Elite system uses RF energy
only for its internal functions. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The Cell Saver Elite system is suitable for
use in all establishments, other than
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC/EN 61000-3-3
Complies