Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
114
Air Detected During Conc
Explanation:
The air detector sensed air in the blue line during the Conc phase. The RBC bag may be empty.
Sequestration has paused.
Corrective Action:
If the RBC bag is not empty:
1. Ensure the blue line clamp is open.
2. Check the tubing for kinks, occlusions, and correct placement.
3. Touch
Continue
Using RBC Bag
to continue.
If the RBC bag is empty:
1. Replace the blood bag.
2. Touch
Continue Using Blood Bag
to proceed.
To end the cycle:
1. Touch
End Cycle
to end the current cycle.
2. The device will empty the bowl contents to the RBC bag.
115
Air Detected Early
Explanation:
The air detector sensed air before the expected volume was pumped from the bowl.
Corrective Action:
1. Check the tubing for correct placement in the air detector and pump.
2. Check the tubing for kinks and occlusions.
3. Touch
Continue
to continue.
116
Air Detector Failure
Explanation:
An issue has been detected with the air detector.
Corrective Action:
Restart the device.
If the problem persists, service is required.