Air Detected During Fill
Explanation:
The air detector sensed air in the red line during the Fill phase. The reservoir may be empty.
Corrective Action:
1. Ensure the red line clamp is open.
2. Check the tubing for kinks, occlusions, and correct placement in the air detector and pump.
3. When additional fluid is in the reservoir, touch
to resume filling the bowl.
OR
Touch
Conc
to continue filling the bowl with red cells from the RBC bag.
OR
Touch
Wash
to wash a partial bowl.
OR
Touch
Return
to return cells to the reservoir.
185
Reservoir Load Cell Calibration Failure
Explanation:
An issue has been detected with the reservoir sensor calibration during the power-on self-test
(POST).
Corrective Action:
Please service the device before continuing use.
186
Waste Bag Load Cell Calibration Failure
Explanation:
An issue has been detected with the waste bag sensor calibration during the power-on self-test
(POST).
Corrective Action:
Please service the device before continuing use.