Chapter 11, Troubleshooting
163
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
104
Power Supply Failure
Explanation:
An issue has been detected with the power supply.
Corrective Action:
Restart the device.
If the problem persists, service is required.
105
Reservoir Sensor Failure
Explanation:
Auto-Fill is disabled. An issue has been detected with the reservoir sensor. Processing can
continue, but the device cannot automatically enter the Fill phase.
Corrective Action:
Touch
Fill
to start Fill when there is adequate fluid in the reservoir.
Touch
OK
to close this message.
Service is required to repair the reservoir sensor.
106
Waste Bag Sensor Failure
Explanation:
An issue has been detected with the waste bag sensor.
Corrective Action:
Restart the device.
If the problem persists, service is required.
107
Sensor Input Failure
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages
ID #
Event Message Text