Chapter 11, Troubleshooting
191
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
233
Bowl Optics Failure
Explanation:
An issue has been detected with the bowl optics. The power-on self-test has paused. The
procedure may be run, but the user must manually start the Wash phase.
Corrective Action:
To proceed manually:
1. Touch
Continue
to continue with POST and run the procedure.
2. Touch
Wash
to start Wash when there is adequate fluid in the bowl.
NOTE:
The bowl is full when the red cell/supernatant interface reaches a point approximately 1/
4 inch over the shoulder of the bowl. The hematocrit of the product may be reduced if the Wash
phase is started before the bowl is full.
OR
Power down the device.
Service is required to fully resolve this issue.
234
Centrifuge Header Arm is Unlatched
Explanation:
The device detected that the centrifuge header arm is unlatched.
Corrective Action:
1. Ensure the bowl is seated properly.
2. Close and latch the centrifuge arm.
3. Close the device cover.
If the problem persists, service is required.
235
Load Cell Calibration Error
Explanation:
An issue has been detected with the calibration data for the load cells. Recalibration of the load
cells is required.
Corrective Action:
Remove the device from service and recalibrate the load cells.
Table 16, Event Messages
ID #
Event Message Text