Chapter 11, Troubleshooting
185
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
183
Effluent Line Sensor Failure
Explanation:
An issue has been detected with the effluent line sensor. The procedure can be performed, but
the device cannot monitor the effluent from the bowl.
Corrective Action:
To perform the procedure with the user monitoring the effluent:
1. Touch
Continue
.
2. Monitor the effluent from the bowl during processing.
If at the end of a wash phase additional washing is required:
1. Touch
Cycle Wash Volume
.
2. Increase the wash volume.
OR
To end the procedure:
1. Touch
End Procedure
.
2. If in the power-on self-test, power down the device.
If the problem persists, service is required to resolve this issue.
Table 16, Event Messages
ID #
Event Message Text