Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
159
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
160
Device Overheated
Explanation:
The internal device temperature may have been high. The red cells might have been damaged
if the parts of the device that contact the processing set rose above 42 °C.
Corrective Action:
1. If the temperature of the device parts that contact the processing set cannot be verified to be
below 42 °C it is recommended to either QC the product prior to reinfusion or to discard the
blood in the processing set and to end the procedure.
2. Please service the device before using it again.
3. Blood remaining in the reservoir may be processed with another device and new processing
set.
161
System Fault
Explanation:
An issue has been detected in one of the internal device components.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages