Chapter 11, Troubleshooting
181
Haemonetics
®
Cell Saver
®
Elite
®
User Manual
P/N 120745-IE, Manual revision: AA
171
Pump Platen Open
Explanation:
The pump platen is not fully closed. The centrifuge must stop before the cover will unlock and
corrective action can be taken.
Corrective Action:
1. Open the pump platen.
2. Ensure correct tubing placement around the pump.
3. Close the pump platen.
4. Close the device cover.
If the problem persists, service is required.
172
Pump Speed Error
Explanation:
The pump is not operating as expected. The centrifuge must stop before the cover will unlock
and corrective action can be taken.
Corrective Action:
1. Open the pump platen.
2. Ensure tubing placement around the pump is correct.
3. Ensure the pump rollers are clean and rotate freely.
4. Close the pump platen.
5. Close the device cover.
If the problem persists, service is required.
173
Suction Failure
Explanation:
An issue has been detected with the internal suction. All other device functions continue to
operate normally.
Corrective Action:
1. Connect an external regulated suction source to the reservoir.
2. Touch
OK
to close this message.
Service is required to repair the internal suction.
Table 16, Event Messages
ID #
Event Message Text